商品編號: EUTECH-alpha-CON1000
alpha-CON1000 Thermo Scientific Conductivity Controller/Transmitter 線上型電導度控制器
   
產品特色
|
 |
液晶螢幕同時顯示電導度, 及溫度值
|
 |
10段測量範圍, 0.000uS...200.0mS
|
 |
4-20mA類比輸出, ON/OFF及警報控制
|
 |
適合超純水, 純水, 印刷水, 農業養殖, 廢水處理...等應用
|
 |
新加坡原裝進口 EUTECH
|
Product Features
The Alpha COND 1000 features up to ten selectable conductivity ranges and corresponding cell constant K values, easily configured through a six-button keypad. Adjustable temperature co-efficient and ultrapure water compensation option allow measurements in broad-range applications, from ultrapure water to high-conductivity samples.
:: |
±1% full scale accuracy across ten different Conductivity ranges |
|
|
:: |
Adjustable temperature coefficient from
0.0 to 10.0% for higher accuracy |
|
|
:: |
Pure water compensation option corrects non-linearity of pure water temperature correction curves in ultrapure water applications |
|
|
:: |
Option of Auto or Manual Temperature Compensation. Three-wire system compensates for cable-length resistance errors |
|
|
:: |
Meter displays electrode information after each successful calibration. Previous calibration data is retained in the event of unsuccessful calibrations |
|
|
:: |
Line adjustment feature corrects long cable resistance errors – an important feature useful in applications involving high conductivity measurements |
|
|
:: |
Galvanically-isolated, scaleable 0/4 to 20 mA output for high-quality output on peripheral devices |
|
|
|
Applications
General: Use for virtually any batch or online applications where rapid, accurate control. Great for OEM/system integrator.
Industrial: Use in applications involving agriculture, chemical processing, boiler and water heaters, wafer-fab, microprocessor manufacturing, pharmaceuticals, pulp and paper industries, and bleach manufacturing.
Water Purification/Treatment: Use to treat batches of incoming process water, ultrapure water, boiler and feed water control.
Regulatory: Hook up to recorder for regulatory compliance data collection.
Specifications
Product Specification
|
Description |
Conductivity |
Range |
... to 1.999 µS/cm ; ... to 19.99 µS/cm ;
... to 199.9 µS/cm ; ... to 1999 µS/cm ;
... to 5000 µS/cm ; ... to 19.99 mS/cm ;
... to 199.9 mS/cm |
Resolution |
0.001 µS/cm ; 0.01 µS/cm ;
0.1 µS/cm ; 1 µS/cm ;
5 µS/cm ; 0.01 mS/cm ;
0.1 mS/cm |
Accuracy |
±1 % full scale reading |
Cell Constant |
0.01 ; 0.1 ; 1 ; 10 |
Temperature |
Range |
-9.9 to 125 ºC |
Resolution |
0.1 ºC |
Accuracy |
±0.5 ºC |
Sensor |
Pt100 / Pt1000 (jumper selectable) |
Compensation |
Auto/manual (normalized at 25 ºC) |
Coefficient |
Ultrapure water or linear 0.00 to 10.00 % |
Set Point and Controller Functions |
Set Point 1 (SP1) /
Set Point 2 (SP2) |
... to 1.999 µS/cm or ... to 19.99 µS/cm or
... to 199.9 µS/cm or ... to 5000 µS/cm or
... to 19.99 mS/cm or ... to 199.9 mS/cm |
Switching Conductivity Hysteresis |
0 to 10 % of full scale |
Function (Switchable) |
P control (pulse length/pulse frequency); limit control |
Adjustable Period With Pulse Length Controller
|
0.5 to 20 sec |
Adjustable Period With Pulse Frequency Controller |
60 to 120 pulse/min |
Pickup/Dropout Delay |
0 to 2000 sec |
Contact Outputs |
3 SPDT relays |
Switching Voltage/
Current/Power |
Max. 250 VAC / max. 3 A / max. 600 VA |
Alarm Functions |
Function (Switchable) |
Steady or fleet (pulse) |
Wash Cycle
|
0.1 to 199.9 hr |
Wash Duration |
1 to 1999 sec |
Pickup Delay |
0 to 2000 sec |
Switching Voltage/
Current/Power |
Max. 250 VAC / max. 3 A / max. 600 VA |
Electrical Data and Connections |
Transmitter function |
0/4 to 20 mA scalable outputs for Conductivity,
galvanically isolated |
Hold function switch |
To freeze output current and deactivate control relays |
Load |
Max. 600 Ohm |
Conductivity input |
2-pin terminal |
Connection terminal |
5-pole, 17-pole terminal, detachable blocks |
Power Supply |
Input |
110 VAC
(jumper selectable) ;
48 to 62 Hz ; max. 7 VA
(For TSCONCTP1001) |
220 VAC
(jumper selectable) ;
48 to 62 Hz ; max. 7 VA
(For TSCONCTP1002) |
Main Fuse |
Slow-blow 250 V / 100 mA |
Pollution Degree |
2 |
Transient Overvoltage Category |
II |
EMC Specifications |
Emitted Interference |
According to EN 50081-1 |
Immunity to Interference |
According to EN 50082-1 |
Environmental Conditions |
Operating temperature range |
-10 to 50 ºC |
Max. relative humidity |
80 % up to 31 ºC decreasing linearly to 50 % at 40 ºC |
Mechanical Specifications |
Dimensions (WxHxD) |
96 x 96 x 175 mm |
Weight |
700 g (unit) / 800 g (packed) |
Ingress protection |
IP54 (front panel) |
Order Information
Order No.
|
Item |
TSCONCTP1001 |
Alpha COND 1000 panel-mount Conductivity controller/transmitter with 110 VAC setting. Incl. terminal blocks, gasket, threaded rods, catch, etc |
TSCONCTP1002 |
Alpha COND 1000 panel-mount Conductivity controller/transmitter with 220 VAC setting. Incl. terminal blocks, gasket, threaded rods, catch, etc |
|
|
KEYWORD: EUTECH, alpha-CON1000, Thermo Scientific Conductivity Controller, Transmitter, 線上型電導度控制器